Product Description
Specification
Measurement Parameter: Oxygen
Measurement Technology: Optical fluorescence quenching
Accuracy: ± (1 % of the reading + 8 ppb)
Response Time: T98 = < 90 s
Primary Media/Application: Pharmaceutical/Life Science Production
ISM (Digital) or Analog: ISM (Digital)
Anti-Bubble Design: No
Wetted Metallic Material: 316L stainless steel
Surface Roughnes of Wetted Material: N5/Ra16 (Ra=0.4μm/16μin)
Sensor Diameter: 12 mm
O-Ring Material: EPDM
Power: 25 V DC 0.1 A
Process Connection: Pg 13.5
Sensor Length: 220 mm
Wetted membrane material: PTFE
Connector Design: Straight
Model Family: InPro 6860i
Autoclavable/Sterilizable: Yes
CIP resistant: Yes
Measuring Temp. Range: 41 °F – 140 °F (5 °C – 60 °C)
Mechanical Temp. Range: -4 °F – 284 °F (-20 °C – 140 °C)
Operating Pressure (Bar): 0.2–6 Bar
Operating Pressure (PSI): 0–87 PSI
Design Pressure: 87 psi (6 bar)
Analog Integration: Simulated electrochemical nA signal
Digital Integration: RS485 - ISM® and MODBUS RTU
Cable Connector: VP6 / VP8 (analog / digital)
Hygienic/Biocompatibility Approvals: FDA/USP Class 7
Hazardous Area Approvals: ATEX/IECEx: II 1/2G Ex ia/ib HC T6 Ga/Gb II1/2D Ex ia/ib nie T83 °C Da/Db
Certificates & Approvals: Quality certificate, Material certificate 3.1, Surface finish certificate, ATEX/IECEx certificate, FDA/USP Class IV
Measurement Range: 0...saturation
Sterilizable/Autoclavable: Yes
Measured Pressure Resistance: 0 psi – 87 psi (0 bar – 6 bar)
Order Guidelines
1. Price & Stock Available on Request. Click to send email to: service@iright.com
2. Please DO NOT make payment before confirmation.
3. Minimum order value of $1,000 USD required.
Collaboration
Tony Tang
Email: Tony.Tang@iright.com
Mobile/WhatsApp/Wechat: +86-17717886924