{"product_id":"mettler-toledo-58044001","title":"Mettler Toledo, 58044001,  Analyzer 3000CS","description":"\n\u003cb\u003eSpecification\u003c\/b\u003e\nChloride Measurement Range: 0-300 ppb\nChloride Limit of Detection: 0.5 ppb\nChloride Measurement Accuracy: ±5% of reading or ± 0.5 ppb whichever is greater\nSulfate Measurement Range: 0-300 ppb\nSulfate Limit of Detection: 1 ppb\nSulfate Measurement Accuracy: ±10% of reading or ±1 ppb whichever is greater\nMeasurement Cycle Time: 45 min typical, programmable between 15 minutes and 1 hour\nSample Flow rate: 25-50 mL\/min\nSample Temperature: 10-45°C (50-113°F)\nSample Pressure: 0.3-7 bar (5-100 psig)\nGrab Sample Measurement: Included 100 mL capacity\nPower Supply: 100-240 VAC, 50-60 Hz, 100 W typical\nDimensions (HxWxD): 927x508x305 mm (36.5\"x20.9\"x12\")\nWeight: 44 kg (97 lbs)\nAmbient Operating Temperature: 10-35°C (50-95°F)\nHumidity: 10-70% non-condensing\nRatings\/Approvals: CE \/ cULus\nProcess Connection: Sample Inlet: 1\/4\" or 6 mm OD tube SS compressin fitting Drain Hose: 19x25.4 mm (3\/4 x 1\") 2 m (6 ft) length included\nOperator Interface: Color touchscreen Simultaneous display of ion concentrations and analyzer status\nAnalog Output: 22mA alarm, 500 ohm max load, not for use with externally powered circuit\nAnalog Output Scaling: Linear, bi-linear, logarithmic (1,2,3,4 decades), auto ranging\nMeasurement Parameter: Chloride and sulfate\nRelay Contacts: Mechanical rated at 250 VAC, 3 Amps (Relay 1 NC, Relay 2 to 4 NO), 4-SPDT Type Reed 250 VAC or DC, 0.5 Amps (Relay 5 to 8)\nAnalog Output Accuracy: ±0.05 mA\nModel Family: 3000CS\nEnclosureType: Full\nBuilt-in sequencer: No","brand":"Mettler Toledo","offers":[{"title":"Default Title","offer_id":47170417393833,"sku":"58044001","price":0.99,"currency_code":"USD","in_stock":true}],"url":"https:\/\/iright.com\/es\/products\/mettler-toledo-58044001","provider":"Iright","version":"1.0","type":"link"}