Product Description
The Sartocheck® 4 plus Bag tester is the result of a clear market request for testing single use bags and bioreactors pre-use. It is based on the same market leading software and hardware platform as the Sartocheck® 4 plus Filter tester and assures full communication compatibility between the both.
As bags require lower test pressures and therefore also higher pressure accuracies than membrane filters, the pressure range has been reduced for an overall optimization of pressure accuracy and pressure drop measurement accuracy. The Sartocheck® 4 plus Bag tester provides the same combination of benefits as the Sartocheck® 4 plus Filter tester:
Barcode Scanner for easy and reliable data entry (optional)
Intelligent selection of test program after scanning the 1D or 2D label
Combination of large, colour touch screen display with keypad
Automated cleaning function eliminates expensive service calls
Multitasking menu
Electronic test reports in PDF format
No thermo paper but dot matrix printer (longer print preservation)
Easy test program and user profile proliferation to other Sartocheck® testers via FTP or SD card
Unparalleled accuracy and repeatability of results for all test types
World class documentation, training, applications, and service support
Allows real time parallel testing by controlling up to four additional test units (optional Bag tester and|or Filter tester MultiUnits)
Fully compliant with 21 CFR Part 11
Developed in accordance with GAMP5
Specifications
Applications:
Market: BioPharma
Connections (Data):
Data Interface: PLC port; FTP; RS232; RS485; RJ45
Dimensions:
Dimensions WxDxH: 460 × 390 × 245 mm
Features:
Languages: English; German; French; Spanish; Italian
Measuring Accuracy - Absolute Deviation Pressure Measurement: max. ±0.3 mbar
Measuring Accuracy - Relative Deviation Pressure Measurement: 0.05%
General Specifications:
Additional Product Specifications: Measuring Ranges - Test pressure: 10 – 300 mbar
Minimum Inlet Pressure: 4 bar g
Pressure Drop: All bag systems up to net volume of 2000 L
Operating Conditions:
Operating Conditions - Temperature: +15 – +35 °C
Operating Pressure: min. 4 bar g; max 10 bar g
Relative Humidity: 10% – 80%
Physicochemical Information:
Pressure Measuring Range: 10 – 300 mbar
Power Requirements:
Input Wattage: 74 Watts
Power Requirements: 100 - 240 V / 47 - 63 Hz
Product Information:
Brand: Sartocheck®
System Type: Integrity Testing Systems
Order Guidelines
1. Price & Stock Available on Request. Click to send email to: service@iright.com
2. Please DO NOT make payment before confirmtaion.
3. Minimum order value of $1,000 USD required.
4. 100% prepayment required.
Collaboration
Tony Tang
Email: Tony.Tang@iright.com
Mobile/WhatsApp/Wechat: +86-17717886924